Method and process for testing the sliding performance of a syringe

Update:31 Oct 2019
Summary:

Whether the sliding performance of the syringe piston i […]

Whether the sliding performance of the syringe piston is appropriate, directly affects the use of clinical medical staff and injection therapy, so the quality of the product is very important, and this indicator is also an important indicator that enterprises must pay attention to.

A mechanical tester is used to move the plunger of the syringe, pumping or draining water while recording the applied force and movement of the mandrel. The tester pulls the syringe plunger at a rate of 100 mm/min ± 5 mm/min and draws water from the sink into the syringe until the baseline reaches the nominal scale capacity line. When the baseline reaches the nominal scale capacity line, stop the core rod movement, zero the recorder again, wait 30s, the tester reverses, push the core rod back to its original position, so that the water in the syringe is discharged into the water tank.

What test instrument is used for the sliding performance test of the syringe? What are the test methods and processes?

Nowadays, the syringe testing instrument industry is developing rapidly, and all kinds of testing instruments are dazzling. Choosing the right high-precision testing instruments can reduce the cumbersome operation process and the test results are accurate and reliable.

The medical syringe sliding performance tester detects the sliding performance of the syringe as follows:

(1) Prepare the sample according to the relevant standard, put the push rod piston into the needle tube, adjust the clamp, and clamp the sample;

(2) In the test type box, select the sliding performance of the piston, and set the corresponding test parameters and sample parameters according to the requirements;

(3) Click to start the test, the sample should not produce leakage; after the single test, the upper chuck returns, after the return position, the result is automatically calculated and returned to the test interface to start the test of the next sample. So after the sample is tested, click on “Generate Results”.

https://www.syringeassemblymachines.com/